Weclearmed® medical gas master alarm panel can be connected to up to 40 zone alarms via the 485 bus. Zone alarm models support the LCD (horizontal) series, LTD series, and LCD (vertical) series.Different series of zone alarms used can be mixed, but the communication address set by the device must be same.
Weclearmed® medical gas master alarm panel main parameters
·Host power supply voltage:DC24V±5%
·Power adapter supply voltage:AC100V~240VAC
·Power Consumption: <10W
·Number of monitoring points: 1~40
·communication method:RS485( Modbus_RTU agreements)
·LCD 尺寸: 10.1 寸( 222.7mm*125.3mm)
·Supports user to define zone alarm names by themselves
·Support sound and light alarm
·Buzzer mute support
·All parameters can be set on-site via touch screen