Weclearmed® medical gas alarm LED compliance with medical gas alarm system EMC electromagnetic compatibility testing(WT18050133) and electrical safety testing(WT18030118). Positive pressure, negative pressure, concentration and flow can be monitored.It support for Mpa, Psi, Bar, mmHg, inHg and other units, to ensure synchronization with other equipment.Wall-mounted design, surface-type and flush-type installation are available.
1. Host power supply voltage : DC9V±5%
2. Power adapter supply voltage : AC100V~240VAC
3. Power Consumption : 2W(1 gas)、3W(2 gases)、4W(3 gases)、5W(4 gases)、6W(5 gases)、7W(6 gases)、8W(7 gases)
4. LED size : 0.8 inch
5. 6. Default pressure range: -0.1MPa ~ 1.0MPa, the number of air pressure monitoring points: 1 ~ 7
6. It has digital real-time display of gas pressure values, local sound and light alarm function.It supports buzzer mute
7. It supports switching output of each air pressure status (instead of electric contact pressure gauge) and DC1~5V analog output of each air pressure value (corresponding to -0.1MPa~1.0MPa)
8. It has RS485 remote communication and standard ModBus protocol.It can be networked and monitored.
9. It support the switching output of the alarm.
10. It supports infrared remote control muting within 5 meters.
11. It support sensor positive and negative pressure common and sensor failure detection.
12. All parameters can be adjusted on site.