1. Comply with EMC test of medical gas alarm system (Wt18050133), and electrical safety test of medical gas alarm system (Wt18030118);
2. positive pressure, negative pressure, concentration, flow can be monitored; Mpa, Kpa, Psi, Bar, mmHg, inHg and other units;
3. 8 Gas alarm program module integrated design, each gas interface can be interchangeable, redundant interface can be used as a standby interface, wall-mounted installation design, make installation and maintenance more convenient.
4. digital real-time display gas pressure value, local acousto-optic alarm and other functions, and provide external remote alarm contact, can achieve remote acoustooptic alarm.
5. central microcomputer processor,Windows operating system, stable and safe, maintenance-free, long service life.
6. standard ModBus protocol,RS485 remote communication, network monitoring
7. the selection of imported components, strong anti-interference ability, all stainless steel high precision pressure sensor, stable and reliable performance.
8. positive pressure sensor and negative pressure sensor can be universal, avoid the occurrence of sensor connection error phenomenon; Supports sensor fault detection
9.secondary password management, all parameters can be debugging, to meet different needs